Description
DF30-G micro syringe pump dispensing system has the characteristics of simple operation, high precision and strong reliability. It is widely used in the filling of various diagnostic reagents, drug research and development, genome and proteomics research.
Function
• 7 inch color touch screen, simple operation, .
• It can support 1-15 channels working at the same time, independent control of each channel.
• With pretreatment functions such as filling and cleaning.
• High repeat accuracy, no need for multiple calibration.
• 5 groups of parameters can be stored and customized for easy to choose.
• No consumables, greatly saving maintenance costs.
• Manual operation (with Footswitch) and supporting automated production line.
Main Interface
Parameters
Filling Volume
|
10μl~5ml(>30% of syringe capacity)
|
Volume Error
|
0.0001~24Erro≤1%(rated stroke)
|
Reliablility Error
|
CV≤0.05%(rated stroke, 25℃ deionized water , 1ml syringe)
|
Rated Stroke
|
30mm
|
Dispensing Time
|
1.2 S~100 Min(rated stroke)
|
Withdraw Time
|
1.2 S~100 Min(rated stroke)
|
Pause Time
|
0.1~999 S
|
Working Times
|
Single running signal
|
Wetted Material
|
Borosilicate glass, PTFE, PCTFE
|
Syringe Fitting
|
1/4-28 inner thread fitting
|
Syringe Size
|
25μl~5ml
|
Running Signal
|
24V level signal input, high level effective(Synchronized control of each channel)
|
Lack bottle stop
|
24V level signal input, high level effective(Individual control of each channel)
|
Dimension
|
540×235×170mm(4 Channels)
|
Shell Material
|
SS304
|
Power supply
|
AC100~240V,50/60Hz,200W
|
Environment
|
Temperature15℃~40℃,Relative humidity<80%
|
Weight
|
10KG(4 Channels)
|
Dispensing Precision ( 25℃, deionized water )
Volume
|
Filling time
|
Withdraw time
|
Volume erro
|
CV
|
Syringe size
|
Diameter mm
|
10μl
|
0.5S
|
0.8S
|
3.0%
|
2.0%
|
25μl
|
1.0×1.6
|
30μl
|
0.5S
|
0.8S
|
2.0%
|
1.0%
|
100μl
|
1.0×1.6
|
50μl
|
0.8S
|
1.0S
|
1.0%
|
0.2%
|
100μl
|
1.0×1.6
|
100μl
|
0.5S
|
1.0S
|
1.0%
|
0.2%
|
250μl
|
1.0×1.6
|
300μl
|
0.5S
|
1.0S
|
1.2%
|
0.5%
|
1000μl
|
1.0×1.6
|
500μl
|
0.8S
|
1.0S
|
1.0%
|
0.1%
|
1000μl
|
1.0×1.6
|
1000μl
|
1S
|
1S
|
1.0%
|
0.1%
|
2.5ml
|
1.5×3.2
|
3000μl
|
2S
|
2S
|
1.0%
|
0.1%
|
5ml
|
1.5×3.2
|
5000μl
|
3S
|
4S
|
1.0%
|
0.05%
|
5ml
|
1.5×3.2
|
Above flow parameters are obtained by using silicone tube to transfer pure water under normal temperature and pressure, in actually it is effected by specific factors such as pressure, medium, etc. Above for reference only.